F317 High Performance Fatigue Rated Cylindrical Loadcell
Standard Ranges 5, 10, 25, 50, 100, 250, 500 and 1000kN (500kgf to 100tonnef)
- Good output symmetry
- High structural load limit
- Fatigue rated
- Sealed to IP65
- Flying lead or connector option
- Traceable calibration with certificate included in the standard price
Specification
Parameter | Value | Unit |
Non-linearity - Terminal | ±0.05 | % RL |
Hysteresis | ±0.1 | % RL |
Creep - 20 minutes | ±0.05 | % AL |
Repeatability | ±0.02 | % RL |
Rated output - Nominal | 1.2 | mV/V |
Rated output - Rationalised | 1.0 | mV/V |
Rationalisation tolerance (applies to single direction calibrations) | ±0.1 | % RL |
Output symmetry | ±0.5 | % AO |
Fatigue life | 108 | RL cycles |
Zero load output | ±4 | % RL |
Temperature effect on rated output per °C | ±0.005 | % AL |
Temperature effect on zero load output per °C | ±0.01 | % RL |
Temperature range - Compensated | -10 to +50 | °C |
Temperature range - Safe | -10 to +80 | °C |
Excitation voltage - Recommended | 10 | V |
Excitation voltage - Maximum | 20 | V |
Bridge resistance | 700 | Ω |
Insulation resistance - Minimum at 50Vdc | 500 | MΩ |
Overload - Safe | 50 | % RL |
Overload - Ultimate | 200 | % RL |
Sealing | IP65 |
All standard ranges are manufactured in stainless steel.
The F317 is a high performance cylindrical loadcell offering a high degree of output symmetry in compression and tension.
A high slenderness ratio strain section improves all round performance compared to the older design F204. New thread sizes are employed offering more economic sizing of rod end bearings and improved availability. Novatech can supply suitable rod ends.
The industry standard fatigue rated loadcell has traditionally been the pancake loadcell, our F254 is a typical example. By restricting the nominal output to a safe strain region the cylindrical loadcell offers the same number of full load reverse cycles as a minimum life expectancy. The cylindrical form offers the advantage of a reduced diameter, often advantageous in actuator applications. Adaptors are not required which are costly and add mechanical interfaces to the load train assembly.
We are happy to design variants of this loadcell to meet your specific requirements. Versions can be manufactured for fully compensated operation up to +250°C. Please consult our engineering department.
Order Codes
Code | Description |
F317CFR0K0 | Compression, IP65, unrationalised |
F317TFR0K0 | Tension, IP65, unrationalised |
F317UFR0K0 | Bi-directional, IP65, unrationalised |
F317CFR0KN | Compression, IP65, rationalised |
F317TFR0KN | Tension, IP65, rationalised |
F317UFR0KN | Bi-directional, IP65, rationalised |
Change the F to a P for the connector version. Not available on the 0.5tonnef range. |
Structural Stiffness - Nominal
Range (kN) | Stiffness (N/m) |
5 | 2.3 x 108 |
10 | 2.5 x 108 |
25 | 6.3 x 108 |
50 | 1.3 x 109 |
100 | 1.4x 109 |
250 | 3.5 x 109 |
500 | 5.6 x 109 |
1000 | 1.2 x 1010 |
Notes
- AL = Applied load.
- RL = Rated load.
- Temperature coefficients apply over the compensated range.
- The load must be applied directly through the central loading axis.
- AO = Average of tension and compression outputs for full load.
Connections
The ranges above 5kN are fitted with 2 metres of PVC insulated 4 core screened cable type 16-2-4C or a 4 pin Binder 723 series chassis plug. The 5kN range is fitted with 2 metres of PVC insulated 4 core screened cable type 7-2-4C, the connector option is not available on this range.
Excitation + = Red or pin 1, Excitation - = Blue or pin 2, Signal + = Yellow or pin 3, Signal - = Green or pin 4, Screen = Orange.
Reverse the signal connections to obtain a positive signal in tension mode. The screen is not connected to the loadcell body.
Files
Type | Title | Download |
Outline | Outline drawing of the product. | |
STEP File | F317-T/C/U-FR0K0 5kN (500kgf) | |
STEP File | F317-T/C/U-FR0K0 10 to 50kN (1 to 5tonnef) | |
STEP File | F317-T/C/U-FR0K0 100 to 250kN (10 to 25tonnef) | |
STEP File | F317-T/C/U-FR0K0 500kN (50tonnef) | |
STEP File | F317-T/C/U-FR0K0 1000kN (100tonnef) | |
STEP File | F317-T/C/U-PR0K0 10 to 50kN (1 to 5tonnef) | |
STEP File | F317-T/C/U-PR0K0 100 to 250kN (10 to 25tonnef) | |
STEP File | F317-T/C/U-PR0K0 500kN (50tonnef) | |
STEP File | F317-T/C/U-PR0K0 1000kN (100tonnef) |