| Novatech

F216 Loadcell

Standard Ranges 5, 10, 20 and 40kN (500kgf to 4tonnef)

  • Hardened stainless steel body
  • Standard 1 year warranty
  • High mechanical stiffness
  • Flying lead
  • Small size
  • Traceable calibration with certificate included in the standard price

Specification

Parameter

Value

Unit

Non-linearity - Terminal

±0.5

% RL

Hysteresis

±0.5

% RL

Creep - 20 minutes

±0.05

% AL

Repeatability

±0.02

% RL

Rated output - Nominal

1.2

mV/V

Rated output - Rationalised

1.0

mV/V

Rationalisation tolerance

±0.5

% RL

Zero load output

±4

% RL

Temperature effect on rated output per °C

±0.005

% AL

Temperature effect on zero load output per °C

±0.03

% RL

Temperature range - Compensated

-10 to +50

°C

Temperature range - Safe

-10 to +80

°C

Excitation voltage - Recommended

10

V

Excitation voltage - Maximum

20

V

Bridge resistance

700

Ω

Insulation resistance - Minimum at 50Vdc

500

MΩ

Overload - Safe

50

% RL

Overload - Ultimate

400

% RL

Sealing

IP65

Weight - Nominal (excluding cable)

90 to 100

g

Geometry: Axial strain cylinder in weather sealed case, with raised end load bearing faces. For use in compression.

The F216 is ideally suited to engineering force measurements. It is designed for easy installation, usually between two flat faces bearing on its loading rings.

The small size of this loadcell makes it an ideal force sensor for mounting in existing machinery. It has been used in presses and printing machines.

We are happy to design variants of this loadcell to meet your specific requirements. Versions can be manufactured for fully compensated operation up to +250°C. Please consult our engineering department.

Order Codes

Code

Description

F216CFR0K0

Compression, IP65, unrationalised

F216CFR0KN

Compression, IP65, rationalised

Structural Stiffness - Nominal

Range (kN)

Stiffness (N/m)

5

2.4 x 108

10

4.8 x 108

20

9.6 x 108

40

1.9 x 109

Notes

  • AL = Applied load.
  • RL = Rated load.
  • Temperature coefficients apply over the compensated range.
  • The load must be applied directly through the central loading axis.

Connections

The loadcell is fitted with 2 metres of PVC insulated 4 core screened cable type 7-2-4C.

Excitation + = Red, Excitation - = Blue, Signal + = Yellow, Signal - = Green, Screen = Orange.

The screen is not connected to the loadcell body.

Files

Type

Title

Download

Outline

Outline drawing of the product.

Download

STEP File

F216CFR0K0 5 to 20kN (500kgf to 2tonnef)

Download

STEP File

F216CFR0K0 40kN (4tonnef)

Download