The F306 is designed for compressive force measurements in situations where space is limited. Alternatively tensile load transfer can be achieved via a tie rod assembly through the centre hole. In this way the loadcell can indirectly measure tensile loads in a ‘fail-safe’ mode.
In the event of structural failure of the loadcell the resulting vertical movement of the supported load will be very small.
We are happy to design variants of this loadcell to meet your specific requirements. Versions can be manufactured for higher temperature operation. Please consult our engineering department.
The F306 is designed for compressive force measurements in situations where space is limited. Alternatively tensile load transfer can be achieved via a tie rod assembly through the centre hole. In this way the loadcell can indirectly measure tensile loads in a ‘fail-safe’ mode.
In the event of structural failure of the loadcell the resulting vertical movement of the supported load will be very small.
We are happy to design variants of this loadcell to meet your specific requirements. Versions can be manufactured for higher temperature operation. Please consult our engineering department.
Parameter | Value | Unit |
---|---|---|
Non-linearity - Terminal | ±0.3 (50N to 1kN) / ±0.5 (2 to 10kN) | % RL |
Hysteresis | ±0.2 (50N to 1kN) / ±0.3 (2 to 10kN) | % RL |
Creep - 20 minutes | ±0.1 (50N to 1kN) / ±0.1 (2 to 10kN) | % AL |
Repeatability | ±0.1 (50N to 1kN) / ±0.1 (2 to 10kN) | % RL |
Rated output - Nominal | 1.2 (50N to 1kN) / 2.2 (2 to 10kN) | mV/V |
Rated output - Rationalised | 1.0 (50N to 1kN) / 2.0 (2 to 10kN) | mV/V |
Rationalisation tolerance | ±0.5 (50N to 1kN) / ±0.5 (2 to 10kN) | % RL |
Zero load output | ±4 (50N to 1kN) / ±4 (2 to 10kN) | % RL |
Temperature effect on rated output per °C | ±0.005 (50N to 1kN) / ±0.005 (2 to 10kN) | % AL |
Temperature effect on zero load output per °C | ±0.01 (50N to 1kN) / ±0.005 (2 to 10kN) | % RL |
Temperature range - Compensated | -10 to +50 (50N to 1kN) / -10 to +50 (2 to 10kN) | °C |
Temperature range - Safe | -10 to +80 (50N to 1kN) / -10 to +80 (2 to 10kN) | °C |
Excitation voltage - Recommended | 10 (50N to 1kN) / 10 (2 to 10kN) | V |
Excitation voltage - Maximum | 10 (50N to 1kN) / 10 (2 to 10kN) | V |
Bridge resistance | 350 (50N to 1kN) / 350 (2 to 10kN) | Ω |
Insulation resistance - Minimum at 50Vdc | 500 (50N to 1kN) / 500 (2 to 10kN) | MΩ |
Overload - Safe | 50 (50N to 1kN) / 50 (2 to 10kN) | % RL |
Overload - Ultimate | 300 (50N to 1kN) / 300 (2 to 10kN) | % RL |
Sealing | IP65 | |
Weight - Nominal (excluding cable) | 20 to 50 | g |
Code | Description |
---|---|
F306CFR0H0 | Compression, IP65, unrationalised |
F306CFR0HN | Compression, IP65, rationalised |
Range (kN) | Stiffness (N/m) |
---|---|
50 (N) | 4.8 x 105 |
200 (N) | 2.2 x 106 |
500 (N) | 8.8 x 106 |
1 | 2.5 x 107 |
2 | 2.4 x 107 |
5 | 9.6 x 107 |
10 | 2.8 x 108 |