Home » Products » Torque Transducers » F315 Low Range Disc Type Static Torque Transducer
This static torque transducer is used extensively in instrumentation and small electric motor test and calibration applications.
Torque transmission must be arranged to input only torque as extraneous forces, such as tare loads or axial loads will cause measurement errors. Note we are able to calibrate with extraneous load cases replicated.
The highly symmetrical signal outputs are usually between 0.2 and 1mV/V for full range.
We are happy to design variants of this transducer to meet your specific requirements. Please consult our engineering department. If you require a higher torque range than 5Nm the F311 is available with ranges up to 2000Nm.
This static torque transducer is used extensively in instrumentation and small electric motor test and calibration applications.
Torque transmission must be arranged to input only torque as extraneous forces, such as tare loads or axial loads will cause measurement errors. Note we are able to calibrate with extraneous load cases replicated.
The highly symmetrical signal outputs are usually between 0.2 and 1mV/V for full range.
We are happy to design variants of this transducer to meet your specific requirements. Please consult our engineering department. If you require a higher torque range than 5Nm the F311 is available with ranges up to 2000Nm.
Parameter | Value | Unit |
---|---|---|
Non-linearity - Terminal | ±0.05 | % RT |
Hysteresis | ±0.1 | % RT |
Creep - 20 minutes | ±0.3 | % AT |
Repeatability | ±0.05 | % RT |
Rated output - Nominal (depends on range) | 0.2 to 1 | mV/V |
Zero load output | ±4 | % RT |
Temperature effect on rated output per °C | ±0.005 | % AT |
Temperature effect on zero load output per °C (1mV/V) | ±0.01 | % RT |
Temperature range - Compensated | -10 to +50 | °C |
Temperature range - Safe | -10 to +80 | °C |
Excitation voltage - Recommended | 10 | V |
Excitation voltage - Maximum | 15 | V |
Bridge resistance | 700 | Ω |
Insulation resistance - Minimum at 50Vdc | 500 | MΩ |
Overload - Safe | 25 | % RT |
Overload - Ultimate | 50 | % RT |
Weight - Nominal (excluding cable) | 110 to 130 | g |
Code | Description |
---|---|
Most F315 loadcells are manufactured to special requirements and are given an F315-Zxxxx number. |